|
|
| |
We support all gages that
output RS232 through a COM Port or Multiplexor.
This section gives additional information about some of the gages supported by
QC-Gage. Limitations, extra features, and unexpected behaviors are documented
here. Exclusion from this list does not indicate a gage is not supported.
Interface notes for more gages, and additional gage definitions may be
available by contacting your sales representative.
Flexbar-Preisser
DIGI-MET
Calipers and indicators with direct RS232 output. These gages are easy
to configure because they have only two modes - setup and measure. In setup
mode you can set a zero point and choose between inches and millimeters.
Connection Notes The default settings are Baud Rate = 4800 baud,
Data Bits = 7, Parity = Even, Flow Control = RTS/CTS (Hardware). Use a straight
cable that passes all lines if you must extend the DIGI-MET's cable. The
DIGI-MET has a female DB9 connector.
If you specify units in the Spec Plan, use in for inches and mm for
millimeters. If you disconnect and reconnect the DIGI-MET from the computer,
the DIGI-MET may send an extra reading or may send characters that block the
first reading, watch the QC-Gage run screen when you connect the DIGI-MET.
Collecting Data To collect data, the DIGI-MET must be in measure
mode, indicated by the words REF 1 in the upper left corner of the display. If
the DIGI-MET is not in measure mode, press and hold button 1 for about 2
seconds and release it. As long as the DIGI-MET is in measure mode you can send
readings to QC-Gage by pressing button 2. The DIGI-MET does not accept the Get
Reading button on the QC-Gage run screen.
Back to Top
GagePort
Gage to RS232 converters that are available for most gages. Configure
the GagePort using switches inside the GagePort and configure the RS232 using
switches inside RS232/power cord connector. Although all GagePorts look alike,
each accepts only a few types of gages. Make sure you have a GagePort that
accepts the type of gage you have, and make sure the switches inside the
GagePort are set to the type of gage for each port. For example, if you had a
GagePort 2220-HH you could set it to accept a Mitutoyo Indicator on Port 1 and
a TESA Caliper on Port 2, but there is no way you could configure it to accept
a Chicago Dial Indicator. You also could not switch the Mitutoyo Indicator to
Port 2 without taking the GagePort apart and changing the switches. The
GagePort has 2 modes - Computer mode and Printer mode. There are 2 separate
gage scripts in QC-Gage for these modes. All ports on the same GagePort must be
in the same mode. In Computer mode the foot switch is disabled and in Printer
mode the Get Reading button is disabled. Note that not all gages respond
to the foot switch and the Get Reading button.
Connection Notes If you use an extension cable to reach the PC,
the cable must contain the hardware handshaking lines. If you are unsure about
your cable, put the GagePort in printer mode, connect the GagePort to the
Microsoft HyperTerminal program, and use the foot switch to send readings. If
HyperTerminal does not receive data, the cable is probably bad. Do not attempt
to connect to QC-Gage until you know the GagePort can send data from the gage
to your computer. If your computer's serial port does not have a a 16550 UART
you should set the baud rate to 1200 baud on both the computer and the
GagePort. Collecting Data Collect data by using
the Get Reading button if the GagePort is in Computer mode, the foot
switch if the GagePort is in Printer mode, or the Send button on the
gage (not all gages have Send buttons).
Back to Top
GagePort NT
A newer version of the popular GagePort Gage to RS232 converters.
GagePort NT converters are available for most gages and include autodetect
capability for many gages. Configure the GagePort NT's gage behavior using a
terminal emulator program like Microsoft's HyperTerminal. Configure the
GagePort NT's RS232 behavior using switches inside RS232/power cord connector.
The GagePort NT has two output modes - Computer mode and Printer mode. Select
the output mode by setting a switch in the RS232/power cord connector. All
ports on the same GagePort will be in the same mode. Use Printer mode with
QC-Gage.
The GagePort NT provides two reading modes, Static and Dynamic. Static mode
takes a single reading and transmits it to QC-Gage. Dynamic mode takes a series
of readings and reports the largest, smallest, and difference between them. You
can have some gages in static mode and some in dynamic mode connected to the
same GagePort NT, but you must be careful how you configure the foot switches
and collect data - it is easy to accidentally set the GagePort to send data
from one gage while it enters dynamic mode for a second gage but this extra
readings from the first gage and the order depends on the gage numbers! If you
use dynamic mode, you must make 3 features in your QC-Gage Spec Plan for each
gage in dynamic mode, even if you don't want all three. The readings come in as
Min, Max, and TIR. If you want QC-Gage to check the mode, specify units of
Static for a feature that should be measured in static mode and units of Min,
Max, and TIR respectively for the three features generated by a reading in
dynamic mode.
The autodetect capability allows quick and easy setup of new gages, and allows
you to change gages with very little effort, but if the gage is not
autodetected or if you need special settings then autodetect may not work well.
The GagePort NT allows you to "burn in" a gage type and settings so when you
power down and power up the GagePort your old settings for that gage come back.
See the "Configuration notes" section for details.
A GagePort with 4 ports can support 4 gages but only has 2 foot switches. You
can configure each port to respond to either, neither, or both foot switches.
This means you can set up a system so pressing foot switch 1 causes gages 1 and
4 to send data, foot switch 2 causes gages 1 and 2 to send data, and gage 3
will only send data when you use the Send button on the gage or press
the Get Reading button on QC-Gage.
Connection Notes WARNING:QC-Gage requires a G-C440 or
G-C441 cable. There are 4 cables that can connect a GagePort NT to a computer -
A-680, A-681, G-C440 and G-C441. The A series cables are used by some GagePort
programs, and are not configurable.
If you use an extension cable to reach the PC, the cable must contain the
hardware handshaking lines. If you are unsure about your cable, put the
GagePort in printer mode, connect the GagePort NT to the Microsoft
HyperTerminal program, set Terminal and the GagePort NT to the same RS232
settings, and use the foot switch to send readings. If Terminal does not
receive data, the cable is probably bad. Do not attempt to connect to QC-Gage
until you know the GagePort can send data from the gage to your computer. If
your computer's serial port does not have a 16550 UART you should set the baud
rate to 1200 baud on both the computer and the GagePort. If you have a 16550
UART you can use 9600 baud, but 1200 baud is better because the volume of data
is small. Most of the processing time is used within QC-Gage, so the slower
data rate increases reliability without affecting the total time to take a
reading. In fact, the GagePort NT has a built in limit of 0.150 seconds between
characters for some functions and sending data faster can cause erratic
behavior in the GagePort! Configuration Notes Configuring
the GagePort NT for the first time can take an hour or more and be somewhat
confusing, but using the configuration procedures provided by Prolink makes the
process much clearer.
Background
The GagePort has three types of memory; "permanent", "semi-permanent", and
"temporary" (actually 4, but we don't need to consider the SmartCables here);
and moving information between them is not always clear. The permanent memory
cannot be changed and contains the default gage definitions, instructions for
autodetection, and instructions that tell the GagePort how to handle
instructions from the computer you connect the GagePort to. All we really care
about this memory is that it contains default gage definitions in a safe place
so if we make a mistake, we can use it to "reset" the GagePort NT to its
original factory fresh state. The second type of memory, semi-permanent, is
readable and writeable nonvolatile memory. This memory is not cleared when you
power down the GagePort NT, so anything put here is permanent until you
explicitly tell the GagePort to change it. When you configure the output
format, foot switches, etc. the settings go here. The last type of memory we
are concerned with is the temporary memory, also known as volatile read / write
memory. This gets reset every time you unplug the GagePort. Readings go here,
and a copy of the gage name if the gage was autodetected. This means that if
you rely on autodetect, the GagePort checks the gage every time you power the
GagePort up so make sure the gage is plugged in! You can copy the gage name
(well, number actually) for a port into the second type of memory, the
nonvolatile read/write memory, so that when you power down and power up the
GagePort NT it uses the one you told it instead of autodetecting. This also
preserves settings like static vs. dynamic mode. You can clear this setting by
specifying a gage type of autodetect (gage number 99) and then burning it in,
but the point is that these settings are taken from the permanent memory by
default but you can change them and store them in the semi-permanent memory if
you want to.
The GagePort NT accepts commands and sends data by RS232. The GagePort is case
sensitive, so the command "As" is not the same as "AS" or "as". It is also
timing sensitive on many commands, and will reject a command sent too quickly
or with too large a delay in it. For this reason we recommend using 1200 baud
with the GagePort NT, even though it can support higher speeds. To prevent
delays in commands you send while configuring the GagePort, you can type the
commands in notepad, then copy and paste them into the terminal emulator you
are using.
Each port on the GagePort can be set or commanded independently. You specify a
port to receive the command by preceding a command with an attention character.
The attention character is different for each port. All attention characters
are upper case letters. The attention character for the first port is A, the
character for the second port is B, and so on. For example, to send the "r"
command (take a reading) to the gage connected to port 3 you would type
Cr
and assuming a gage was connected to port 3, was turned on, and was in static
mode you would get a reading back within a second or two. Most of the examples
here use port 1 (attention character A) but you could apply them to ports 2, 3,
etc. by replacing the A with a B, C, etc.
How to start
The first step is to set the RS232 parameters. Take the shell off the
RS232/power connector for the GagePort. Inside the shell are DIP switches for
printer vs. computer mode, baud rate, and parity. Set the switches for printer
mode, CRLF, 1200 baud, no parity. You can use 9600 baud, but using 1200 assures
you don't overrun the GagePort's input buffer and so little data passes at a
time that the total transfer time during data collection differs from 9600 baud
by less than 5%.The next step is to connect the gages to the GagePort and the
serial / power cable between the GagePort and the computer and plug the power
cable in. Start Microsoft's HyperTerminal or the terminal emulator of your
choice. Choose the serial port you connected the GagePort NT to, set its baud
rate to 1200, set parity to none, data bits to 8, stop bits to 1, and flow
control to none. Close and reopen the serial port to make sure the new settings
are in effect. You may also want to save these settings so you can easily
reconfigure the GagePort in the future. You should now be able to type
Ai
and get a response similar to the following one
iP01,M2104,2.4,U0D0,G#099(051),f=1-2-e-,o=P+U+N+H-
the exact response is not important, but there should be no strange characters
in it.
If you did not get a response, or the response is garbled, you have a problem
with your settings, cable, computer, or GagePort and cannot proceed until it is
fixed. The QC-Gage manual contains an appendix with a section on debugging
cable problems, and the GagePort manual describes the cable switch settings.
If your GagePort has been used and possible configured before, you should reset
it to factory fresh condition by disconnecting it from the serial / power
cable, removing the cover, putting the "reset jumper" in place, and
reconnecting the serial / power cable. The GagePort NT will display the
following menu
Emulating 2104 Number of ports=4
H=Hexdump
P=Poke addr
K=checK_sum calculation
R=Reset and run
Type R to reset it to factory fresh condition, wait about 10 seconds, and then
disconnect the serial / power cable. Remove the jumper, put the cover back on
the GagePort, and attach the serial / power cable again.
What to set
Remember that typing too slow, or forgetting to press Enter after typing a
command can cause the GagePort to reject the command!
Disable the transmission of headers on power up with the following line. You do
not need to repeat it for each port because all ports share the same output
format. You do not need to "save" this setting, it automatically goes into
semi-permanent memory.
Aoh-
(optional)Set the foot switch action for each port. The format is Aw followed
by the foot switches, if any, you want port A to respond to. If you want port A
to respond to both, use 1,2 after the w. Repeat with ports B, C, and D if
desired. You do not need to "save" foot switch settings, they automatically go
into semi-permanent memory. To make foot switch 1 cause gages 1 and 4 to send
data, foot switch 2 cause gages 1 and 2 to send data, and gage 3 ignore the
foot switches you would send the following lines to the GagePort.
Aw1,2
Bw2
Cw
Dw1
(optional)Force power on gage types for each channel. Use the d command to set
the port to a specific gage and the b command to burn the setting into
semi-permanent memory. The d command also lets you set the user bit (used by
the GagePort for alternate modes in some gages) and dynamic / static mode. The
format is
Adxy#zzz
where x is the user bit (usually 0), y is 0 for static or 1 for dynamic, and
zzz is the gage number. Use a gage number of 99 to go back to autodetect mode.
For example, to set port 1 to a Federal Maxim in static mode, port 2 to a
Federal Maxim in dynamic mode, port 3 to autodetect a gage and use dynamic
mode, and port 4 to a UltraDigit III in static mode use the following 4 lines.
Ad00#62
Bd01#62
Cd01#99
Dd00#60
(optional)Save your gage type and reading mode settings with the b command. You
must repeat it for each port you want to save. To save ports 1, 2, and 4 use
the following lines
Ab
Bb
Db
Checking your settings
When you have made all the settings you need, you should check them to make
sure they were accepted. Timing errors or typing errors can result in the
GagePort NT rejecting a command you thought was accepted. Type
Ai
Bi
etc. for each port and compare the settings with what you expected. The return
from an "i" command is a string in the form
Paa,UbDc,G#099(ggg),f=1h2iej,o=PkUmNnHp
or
Paa,UbDc,G#ddd,f=1h2iej,o=PkUmNnHp
where aa is the port number (01 to 04 for a 4 port GagePort), b is the user
bit, c is 0 for static or 1 for dynamic mode, ddd is the gage number (099 for
autodetect), ggg is the autodetected gage number, h is + for respond to foot
switch 1 or - for ignore it, i is + for respond to foot switch 2 or - for
ignore it, j is + for send an event or - for no events, k is + for include port
number in output or - for do not include, m is + for include register name in
output or - for do not include, n is + for include reading number in output or
- for do not include, p is + for transmit header information on power up or -
for do not transmit. For example
P01,U0D0,G#099(051),f=1-2-e-,o=P+U+N+H-
QC-Gage requires settings of - for j and p, and + for k, m, and n. Settings for
a, b, c, d, g, h, and i will affect how the GagePort interacts with the gage
and with the foot switches, but not how it interacts with QC-Gage.
Collecting data In static mode collect data by using the Get
Reading button, the foot switch, or the Send button on the gage
(not all gages have Send buttons).
In dynamic mode, collect data by pressing the foot switch once to start taking
readings, then pressing it again to end the dynamic reading cycle and transmit
the results. Note that for the GagePort NT in dynamic mode the Get Reading
button is not disabled but does not work as you might expect - it starts
dynamic reading but you must use the foot switch to end and transmit the
reading!
If a gage gets disconnected from the GagePort while the GagePort is collecting
data, a dynamic mode reading collects more than 9999 readings, or the GagePort
is asked to collect data from a port that does not have a gage connected to it
the GagePort NT may stop responding to foot switches and / or the computer. If
this happens, disconnect the serial / power cable from the GagePort and
reconnect it. This clears the error and restarts the GagePort.
Back to Top
GageMux
Gage to RS232 converter made by Gage Connections. Configure the GageMux
using DIP switches. The GageMux is available in 2 and 4 port models and
supports a wide variety of gages.
Connection Notes The default settings are Baud Rate = 9600 baud,
Data Bits = 8, Parity = None, Flow Control = None. Use a straight through cable
to connect the computer to the GageMux. When you write a Spec Plan, make sure
you either leave the Units field blank or put the gage mode in the Units
field. The modes are NRM, MIN, MAX, and TIR. Collecting Data
Collect data by using the Get button in QC-Gage or the foot switch on
the GageMux. Back to Top
Metronics QC-1000 and
QC-1100 For interface to QC-Gage the QC-1000
and QC-1100 are identical so anywhere you see QC-1000 you can read it as
QC-1100. The Quick-Chek is a digital read out with RS232 output capability. You
can enable or disable each axis independently so make sure your Spec Plans
include enough features for all readings. For example, if you have X, Y, and Q
axes enabled, each time the operator pushes the print button the QC-1000 will
fill 3 features in QC-Gage. The QC-1000 reports the following units of measure:
inch, mm, deg, and dms for inches, millimeters, decimal degrees, and degrees
minutes seconds respectively. When you write Spec Plans you can leave the Units
field blank or use one of the abbreviations listed above.
Connection Notes The default settings are Baud Rate = 600 baud,
Data Bits = 8, Stop Bits = 2, Parity = None, Flow Control = Hardware. Use a
custom cable to connect the computer to the QC-1000. The cable must have a 9
pin male connector on the QC-1000 end and a 9 pin female or 25 pin female
connector on the computer end.
|
Cable Wiring Diagram for QC-1000 to PC |
| QC-1100 (9 pin male) |
PC (9 pin female) |
PC (25 pin female) |
| 1 |
5 |
7 |
| 4 |
2 |
3 |
| 5 |
8 |
5 |
| 7 |
7 |
4 |
| 8 |
3 |
2 |
Collecting Data Collect data by using the Get
button in QC-Gage, the foot switch on the QC-1000, or the print button on the
QC-1000. If you have a probe attached to the QC-1000 and it is setup correctly
(see the QC-1000 manual) the probe can trigger data collection also.
Back to Top
Metronics QC-2000
The Quadra-Chek 2000 is used in many inspection devices, marketed under
many different names. It can be run in immediate mode or programmed. Programmed
mode works better with QC-Gage because the QC-2000 will send exactly the same
features in exactly the same order. In immediate mode the operator must be more
careful because if he changes the way he prints data or the order he measures a
feature he can put the data in QC-Gage in the wrong order. The QC-2000 has
several different output modes and formats. The Gage Script we provide works
with 32 column format, either printer mode or computer mode.
There is a difference between the QC-2000 and the QC-Gage definitions of the
word "Feature". The QC-2000 uses this word to describe an entity on a part,
like a circle or a line. QC-Gage uses the word to describe a single
measurement, like the diameter of a circle or the X position of a line. A
single QC-2000 feature may have several QC-Gage features. Although the QC-2000
manual does not have a term for the individual measurements, you can think of
them as attributes. Using this terminology, QC-Gage features are attributes of
a QC-2000 feature.
You can send data from the QC-2000 three ways - you can send specific
attributes of a feature using the ./X, 0/Y, +-/XY and More
buttons, you can send all data for the current feature with the Print button,
or you can send all data for all features with the "Print all features"
function. Sending specific attributes of a feature works best.
Sending data using the ./X, 0/Y, +-/XY and More buttons
allows you to send exactly what you want. You simply click the More button
until the attributes you want, i.e. X and Y position or +T and -T, are
displayed in the Message Display Window and then you press the ./X,
0/Y, or +-/XY button. For example, if you wanted to send a
circle's radius and its X and Y position, you would probe the circle and press
the Finish button. The Message Display Window would show r/d and
their values would be in the X display and Y display. You want the radius so
press the ./X button to send it. Next press the More button, the Message
Display Window now shows X/Y. You want both so press the +-/XY button.
NOTE: If you look at the data (using Microsoft's Terminal.exe or using a
printer) you will see all data labeled X or Y. In our example you would see
three lines of output - an X (the radius) a second X (the X position) and a Y
(the Y position). QC-Gage replaces the labels the QC-2000 sends with the labels
you provide when you write a Spec Plan so you don't need to worry about these
labels, but you should be aware of this behavior in case you are trouble
shooting your program using a printer or Terminal.
Sending data using the Print button (or the "Print all features"
function) on the QC-2000 produces a list of attributes for the current feature
(or all features), including "form data" (a +T and a -T reading) for most
features. See the QC-2000 manual for a description of form data. QC-Gage is set
to ignore the +T and the -T for all features. If you need form data, you must
send it by pressing the More button until the Message Display Window
shows +/-T and then pressing the ./X, 0/Y, or +-/XY button
to send the data.
When you output data using the Print button or the "Print all features"
function the QC-2000 sends multiple measurements on one or more features to
QC-Gage. If you will use either of these methods to output results you must
make sure your Spec Plans contain enough room for all the measurements, even if
you don't want them all. For example, if you want to measure 2 circles, your
Spec Plan must have 8 features in it - an X position, a Y position, a Radius,
and a Diameter for each circle. The presence of these extra measurements is the
reason most people prefer using the ./X, 0/Y, +-/XY and More
buttons to send data.
When you write Spec Plans in QC-Gage, make sure the units match the units the
QC-2000 is set for. The abbreviations the QC-2000 uses for units are "in" and
"mm" for inches and millimeters respectively. Angles do not report units, all
angles are measured in degrees.
The QC-2000 should have the following settings, accessed by pressing the 5/Setup
key, pressing Enter/Yes to enter setup, and using the 8/UpArrow, 2/DownArrow
and Enter/Yes keys to select menu choices. See the QC-2000 manual for
details. Set Angle Types to DD Angles. Set Feature Format, Form
Feed to no. Set Feature Format, 32 Column to yes
(checked). Set RS232, POE to no. In RS232, set 1st,
2nd, and 3rd for Pre Line, Pre Form, and Post Form
to nothing (displays as a ?). In RS232, Post Line set 1st
to 13 and set 2nd and 3rd to nothing.
QC-Gage has two similar scripts for the QC-2000, the "QC-2000 with Init" and
the "QC-2000 32 col mode". The scripts are identical except the "QC-2000 with
Init" script initializes the QC-2000 to settings QC-Gage requires. The settings
made are 32 column reports, angle type to DD, no form feed, and no POE. It does
not set pre and post line and form characters (see the QC-2000 manual), nor
does it set RS232 settings or any toggles (like polar / Cartesian or inch /
mm). The "QC-2000 with Init" script starts slower because it takes time to send
the commands to the QC-2000. The QC-2000 must be turned on and connected to the
computer before you run a QC-Gage Spec Plan that uses a "QC-2000 with Init"
script. Please note that most users will set the QC-2000 up with the settings
QC-Gage needs and not change it, you only need the "QC-2000 with Init" script
if you frequently change between QC-Gage and another form of output.
Connection Notes Use a straight cable with a male DB25 connector
on the QC-2000 end and a Female DB9 or DB25 cable on the computer end. The
cable should pass hardware handshaking lines because "Printer mode" on the
QC-2000 uses hardware flow control. If you do not have a cable that passes
hardware handshaking lines, you must use "Computer mode".
If your computer's serial port does not have a 16550 UART you should set the
baud rate to 2400 baud on both the computer and the QC-2000. The default
settings are Baud Rate = 4800 baud, Data Bits = 8, Parity = None, Stop Bits = 2
Flow Control = Software (XOn / XOff). Collecting Data Collect
data by using the More button to select the attributes you want and the ./X,
0/Y, or +-/XY buttons to send them. Alternately, you can use the Print
button, the "Print all features" function, see the description above.
Back to Top
Mettler
Toledo A wide variety of instruments for
measuring weight, moisture content, pH, temperature, etc. Most support RS232
output in several modes. Some instruments cannot support both RS232 and
LocalCAN, you must choose one or the other.
Connection Notes Use a cable that passes handshaking lines.
Collecting Data Make sure the instrument is in single reading
mode, not in statistics, identification, or calibration mode. Collect data by
using the Print button or the external foot switch on the instrument.
Back to Top
Micro-Vu Q16
An optical comparator with RS232 output. The Q16 sends all available
output (usually 3 or 4 readings, depending on the mode) so make sure you put
enough features in your Spec Plans. For example, if the Q16 is in angle mode
you get 4 readings every time you push the XMIT button - X, Y, Screen Rotation,
and Angle.
Connection Notes The default settings are Baud Rate = 2400 baud,
Bits per Byte = 8, Stop Bits = 1, Parity = None, Flow Control = Hardware. You
can change the baud rate by removing the switch cover on the back panel of the
Q16 and setting the switches according to the Q16 manual, but do not use baud
rates above 2400 unless your computer has 16550 compatible UART in the serial
port. Connect the Q16 to the computer with a straight cable that passes
hardware handshaking lines. The cable should have a 25 pin connector on the Q16
end and a 9 pin female or 25 pin female connector on the computer end. Use the
XMIT port not the PRINT port. Collecting Data Collect
data by using the Get Reading button on the QC-Gage Run Screen or the
XMIT button on the Q16. Back to Top
Mitutoyo DMX-2
A Gage to RS232 converter used with profile projectors, etc. The DMX
normally transmits both channels (X and Y) simultaneously. You should either
make spec plans that collect readings in pairs, or use the Get Reading button
to collect one channel at a time.
Connection Notes Use a straight cable with a 25 pin male
connector on the DMX end. By default the QC-Gage script uses 9600 Baud (with no
parity, 8 data bits, 1 stop bit) and no handshaking, but the DMX does
understand hardware handshaking. If you use handshaking, the cable must contain
the hardware handshaking lines. If you are unsure about your cable, connect the
DMX to the Microsoft Terminal program, set Terminal to use hardware
handshaking, and send 1[CR] (where [CR] means press the enter key) to the DMX.
If Terminal does not receive data, the cable is probably not right for the DMX.
Note that baud rates higher than 2400 baud require a serial port based on a
16550 UART. Collecting Data Collect data by
using the Get Reading button or the foot switch. If you use the foot
switch the DMX sends both channels at the same time. If you use the Get Reading
button the DMX sends only the reading for the channel specified for the current
feature. Back to Top
Mitutoyo MRS100
Gage to RS232 converter that connects a Mitutoyo gage to a serial port.
Connection Notes The default settings are Baud Rate = 1200 baud,
Data Bits = 8, Stop Bits = 1, Parity = None, Flow Control = None. The MRS
includes a cable with a DB25 female connector that connects directly to the
computer. If you need to extend the cable use a straight through cable that
passes all lines. If your computer has a 9 pin serial port use a 25 to 9 pin
adapter that is male on the 25 pin side and female on the 9 pin side.
Collecting Data Collect data by using the Send button on
the gage. Back to Top
Mitutoyo Mux10
Gage to RS232 converter that connects up to 3 Mitutoyo gages to a
single serial port. Configure the Mux10 using switches inside the Mux10.
Connection Notes The default settings are Baud Rate = 1200 baud,
Data Bits = 8, Parity = None, Flow Control = None. Use a straight through cable
to connect the computer to the Mux10. The Mux10 end of the cable is a male DB25
connector. Collecting Data Collect data by
using the Get Reading button in QC-Gage, the Load switch on the
Mux10, or the external foot switch on the Mux10.
Back to Top
Qualitron GageNET
Stackable interface boxes that multiplex dozens gages onto a single
RS232 port. Static or dynamic mode (MIN, MAX and TIR) can be set for each gage
individually. Configure the GageNET using switches on the back of the GageNET.
Note that not all gages respond to the foot switch and the Get Reading button.
When you write Spec Plans, you must use 3 features for each measurement
made in dynamic mode. Each reading in dynamic mode includes MIN, MAX and TIR
values, in that order. Make sure the features are in order and all have the
correct channel number. You can use the units field in the spec plan (go
to Edit - Spec Plan Features) to check for the correct mode. If
you put in units of NRM for readings that should be taken in static mode and
MIN, MAX, and TIR for units of the first, second, and third feature of each
reading in dynamic mode, QC-Gage will warn the operator if the gage is in the
wrong mode. If you use any other text for units, mm for example, QC-Gage does
not check the mode. Connection Notes Use a
cable with a male DB25 connector on the GageNET and a Female DB9 or DB25 cable
on the computer end. If your computer's serial port does not have a 16550 UART
you should set the baud rate to 2400 baud on both the computer and the GageNET.
The default settings are Baud Rate = 9600 baud, Data Bits = 8, Parity = None,
Stop Bits = 1 Flow Control = None. Collecting Data Collect
data by using the Get Reading button on the QC-Gage Run Screen, the foot
switch on the GageNET, or the Send button on the gage (not all gages
have Send buttons). Back to Top
Raw Data
Use this gage definition for any gage or device that transmits a
single, unformatted reading with only a carriage return. The data must be in
ASCII format. For example, -4.454
6500
.0003
All are valid lines but
4.5 mm
12, 4.53
3.55, 3.51, 3.54
are not valid lines. You won't get any warnings when
you use this gage definition because it does not check for channel, units, etc.
Connection Notes Your connection will depend on the gage you
use, refer to its manual. Collecting Data The Get
Reading button on the QC-Gage Run Screen will not work with this gage
definition, see the manual for the gage you are using for instructions on
sending data. Back to Top
Wilson - 500, 600
Hardness testers built by Wilson / Instron. The Wilson 2000 can also
emulate the 500 or the 600. The Wilson 2000 emulation does not exactly match
the true Wilson 600 format, so use the script "Wilson 600" for a Wilson 600 and
"Wilson 600 Emulation" for a Wilson 2000 in 600 emulation mode (and possibly
for newer Wilson 600 testers). The "Default" and "Main and Final Result"
formats of the Wilson 2000 can detect incorrect units and are preferred.
Back to Top
Wilson - 2000
A hardness tester built by Wilson / Instron. This gage has 4 output
formats, each has a specific "Gage Type" setting in QC-Gage. The output formats
are "Default", "Wilson 500 Emulation", "Wilson 600 Emulation", and "Main and
Final Result". Select the output format in the Setup menu Printing
submenu Result Format option. We strongly recommend the "Default"
format.
Note that not all menu choices are on the screen at the same time. You must use
the up and down arrows to page through the menu choices. If there are 6 choices
in a menu, 5 will be on one page and 1 choice will be alone on a second page.
Connection Notes You can configure the Wilson 2000 to use a
variety of serial port settings. Use the Setup menu RS232 option
and pick the setting you want to change. You can set the baud rate to 2400 baud
and use a computer without a 16550 UART based serial port in most cases. The
default settings are Baud Rate = 2400 baud, Data Bits = 8, Parity = None, Flow
Control = RTS/CTS (Hardware). Use a null modem cable to connect the computer to
the Wilson 2000. The Wilson 2000 RS232 connector is a male DB9 connector.
Make sure you enter the units correctly in the Spec Plan, QC-Gage gives a
warning message if the units don't match. QC-Gage also gives a warning if you
are measuring in one scale and converting to another scale but the reading is
not convertible (indicated by a --.- on the 2000 screen). Collecting
Data Collect data by leaving the "Auto Print" mode of the 2000
on, or by pushing the Print button followed by the Clear Stats button
after each set of measurements. If you forget to press the Clear Stats button
or if you use Auto Print and press Print, you will get some
readings twice. Back to Top
Wilson - Tukon
A hardness tester built by Wilson / Instron. This gage has 4 output
modes, only one of which is supported by QC-Gage.
Warning The Tukon system will not transmit any data if QC-Gage
is not up and running when the gage is turned on. If Tukon does not "see" a
computer or printer connected to the gage it refuses to send data. Be sure to
turn the equipment on in the correct order - PC and QC-Gage first, then Tukon!
QC-Gage ignores all data sent to it except Format 3. This yields 3 or 4 columns
of numbers depending if Knoop or Vickers scales are used. You can write a Spec
Plan to ignore any of the undesired values but you must include the correct
units. In the case above either uM, HK or HRC. If these units are not correctly
entered into the Spec Plan, the data is ignored.
Connection Notes The Tukon requires the following serial port
settings - Baud Rate = 2400 baud, Data Bits = 8, Parity = None, Flow Control =
None. Collecting Data There are 4 buttons that
send information through the serial port of the Tukon to an attached printer or
in this case, to QC-Gage. The button that sends it in the correct format is the
trigger on the handle (FILAR). Format 1 Pushing the Print button
on the Digital FILAR Display sends the display reading.
Format 2 Pushing the Stat button on the Digital FILAR Display sends a
summary.
Format 3 Pushing the Trigger on the Digital FILAR Display sends the
reading in several scales. This is the format QC-Gage needs.
Format 4 Pushing the Stat button on the back of the console sends raw
data and a stat summary.
Back to Top |
|